5. Operating the Instrument
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
103
5.8
Multiple Reads per Well
The i-control software allows Multiple Reads per Well (MRW) to be performed in
absorbance, fluorescence top and fluorescence bottom mode.
The Multiple Reads per Well functions can be activated on an absorbance or
fluorescence intensity program strip by selecting the ‘Multiple Reads per Well’
check box (see Figure 40below).
Figure 40: Multiple Reads per Well
Note:
The function ‘Multiple Reads per Well’ is only available for the fixed
wavelength reading modes ‘absorbance’, ‘fluorescence intensity top’
and ‘fluorescence intensity bottom’. The function is not available for
scan measurements.
5.8.1
MRW Type
The MRW types define the pattern how the measurement will be performed. The
software allows seven different MRW types to be selected:
•
Square
•
Square (filled)
•
Circle
•
Circle (filled)
•
X-line
•
Y-line
•
XY-line