3. Installation
48
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
3.1.1
Unpacking Procedure
1. Visually inspect the container for damage before it is opened.
Report any damage immediately
.
2. Select a location to place the instrument that is flat, level, vibration free,
away from direct sunlight, and free from dust, solvents and acid vapors.
Allow at least 10 cm distance between the back of the instrument and the
wall or any other equipment. Ensure that the plate carrier cannot be
accidentally hit when moved out. Ensure that the main switch and the
main cable can be reached at all times and are in no way obstructed.
3. Place the carton in an upright position and open it.
4. Lift the instrument out of the carton and place it in the selected location.
Take care when lifting the instrument and ensure that it is held on both
sides.
5. Visually inspect the instrument for loose, bent or broken parts.
Report any damage immediately.
6. Compare the serial number on the rear panel of the instrument with the
serial number on the packing slip.
Report any discrepancy immediately.
7. Check the instrument accessories against the packing list.
8. Save packing materials and transport locks (see next section) for further
transportation purposes.
WARNING
THE
INFINITE 200 PRO
IS A PRECISION INSTRUMENT AND
WEIGHS FULLY EQUIPPED APPROX. 16 KG.
CAUTION
THE MAXIMUM LOAD FOR THE
INFINITE 200 PRO
COVER IS 16 KG;
HOWEVER THE LOAD MUST BE DISTRIBUTED EVENLY ACROSS THE
ENTIRE SURFACE OF THE COVER.
CAUTION
THE MAXIMUM LOAD FOR THE
INFINITE 200 PRO
PLATE
TRANSPORT IS 100 G. OVERLOADING THE PLATE CARRIER CAN
CAUSE INSTRUMENT DAMAGE WHICH MAY REQUIRE SERVICE.
CAUTION
ALLOW AT LEAST 10 CM DISTANCE BETWEEN THE BACK OF THE
INSTRUMENT AND THE WALL OR ANY OTHER EQUIPMENT.
CAUTION
THE INSTRUMENT MUST BE PLACED IN A LOCATION AWAY FROM
DIRECT SUNLIGHT. ILLUMINATION > 500 LUX CAN NEGATIVELY
INFLUENCE LUMINESCENCE MEASUREMENTS.