5. Operating the Instrument
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
117
Empty Waste Tub Procedure (i-control):
Click the
‘Empty Waste tub’
button and the plate carrier will move out
automatically. Remove the waste tub and empty the contents. After the waste tub
has been emptied place it back on the plate carrier. The i-control software will
alert you when the waste tub needs to be emptied again.
CAUTION
PLACE THE WASTE TUB ON THE PLATE TRANSPORT BEFORE
STARTING A MEASUREMENT WITH THE ACTIONS ‘INJECTION’
AND/OR ‘DISPENSE’.
CAUTION
IT IS RECOMMENDED TO EMPTY THE WASTE TUB BEFORE
STARTING A MEASUREMENT AND TO EMPTY IT AT LEAST ONCE A
DAY.
WARNING
BIOLOGICAL HAZARDS CAN BE ASSOCIATED WITH THE WASTE
MATERIAL (MICROPLATE) OF THE PROCESSES
RUN ON THE
INFINITE 200 PRO
.
TREAT THE USED MICROPLATE, OTHER DISPOSABLES, AND ALL
SUBSTANCES USED, IN ACCORDANCE WITH GOOD
LABORATORY PRACTICE GUIDELINES.
INQUIRE ABOUT APPROPRIATE COLLECTING POINTS AND
APPROVED METHODS OF DISPOSAL IN YOUR COUNTRY, STATE
OR REGION.
STOP
STOP