7. Quality Control
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
171
7.3.8
Absorbance Baseline Flatness (Infinite M200 PRO)
Perform the following measurement to determine the baseline-flatness:
Measurement Parameters:
Parameter
Setting
Reading Mode
Absorbance
Measurement Wavelength from/to
300 - 700 nm
Number of flashes
25
Settle Time
0
Plate Type
GRE96ft
Part of the Plate
A1
Plate Layout:
No plate is necessary for measurement – the plate carrier has to be empty for this
measurement.
Material/Reagents:
No material or reagents necessary for this test.
Calculation of Baseline Flatness:
Calculate the standard deviation.