3. Installation
50
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
CAUTION
SAVE PACKING MATERIALS AND TRANSPORT LOCKS FOR FURTHER
TRANSPORTATION PURPOSES. THE
INFINITE 200 PRO
MUST BE
SHIPPED ONLY WITH THE ORIGINAL PACKAGING AND INSTALLED
TRANSPORT LOCKS.
3.3
Transport and Storage
3.3.1
Transport
The Infinite 200 PRO must be shipped using the original packing and installed
transport locks. Before shipping the instrument, it must be thoroughly disinfected
(see 8.4 Instrument Disinfection).
3.3.2
Storage
Before storing the instrument the injectors must be rinsed using a wash
procedure (see 5.10.2 Priming and Washing of the Infinite 200 PRO). Select a
location to store the instrument that is flat, level, vibration free, away from direct
sunlight, and free from dust, solvents and acid vapors
Storage Specifications
Temperature
- 20 °C to + 60 °C
-4 °F to + 140 °F
Relative Humidity
< 80 % non condensing
STOP