7. Quality Control
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
173
7.3.10 Absorbance Cuvette (Infinite M200 PRO only)
For the Infinite M200 PRO with the option ‘Cuvette’ the following test may be
performed to prove the accuracy specification:
Accuracy
Perform the following measurement to determine the absorbance accuracy:
Measurement Parameters:
Parameter
Setting
Reading Mode
Absorbance
Measurement Wavelength
1) 440 nm
2) 635 nm
Number of flashes
25
Time between Move and Read
0
Cuvette Type
Calibrated cuvette, e.g. Starna
RM- a D3 Cuvette
Material:
Starna
®
reference material RM- D3 cuvette
(for more information please refer to www.starna.co.uk)
Calculation of Accuracy:
Calculate the deviation of the measured value from the reference value supplied
with the calibrated cuvette.