12-4
Color Unevenness Inspection
12
2) On the "MEA-CND" setting screen, press the
TRG/BRT key to move the cursor to the upper
function menu. Select the "COND" item and
then select the "REG COND" item and press
the SET key.
3) On the menu tree, move the cursor to the
"MEAS CND" item, and press the right arrow
key. The sub menu which contains the "REG
COND" item will appear. Move the cursor to
this item and press the SET key.
Set the register conditions
REGISTER NO.
The currently selected register number is
displayed.
If you want to set the measurement conditions
for a different register number, press the SET
key to highlight the register number currently
selected. Then press the up and down arrow
keys to select the register number whose
measurement conditions you want to set.
Press the SET key to confirm your selection.
Note: When the "BINARY-IMG-MASK" or the
"POLYGON" is selected at
"
WINDOW" on the "MEA-CND"
screen, the "
REGISTER NO." line
will not be displayed and the other item
numbers are decrement by one since
the register number for "BINARY-IMG-
MASK" and "POLYGON" is fixed.
REFERENCE COLOR
Select a reference color for color extraction
from the already registered colors (when
more than two colors are registered).
MEAS SHAPE
Specify shape of measuring range.
OBJECT TYPE COND
TYPE00
TYPE RUN COND
IMAGE-ADJ
MEA-CND(CAMERA1)
MEAS0 (POSI-DEVIATION F)
POSI-CORRECT
MEAS01 (
INSPECT COLOR SHADING
P)
MEAS CND
REG COND
EVALUATION COND
NUM-CALC
OUT-COND
MEAS(NEW)
MEA-CND(CAMERA2)
FINAL NUM. CALC
FINAL OUTPUT COND
OBJ-TYPE I/O
OBJ-TYPE SYS.
1
REGISTER NO.
2
REFFERENCE COLOR
3
MEAS SHAPE
4
MEAS AREA
5
BINRY NOISE FILT
6
NUM.OF FILT PASS
COLOR F C1
BRT
SELECT OBJECT TYPE COND
SET
=TO NEXT SUB-MENU ESC=BACK SEL=CHNG IMG TRG=POPUP
+
1
REGISTER NO.
0 (0
〜
3)
2
REFERENCE COLOR
0 (0
〜
7)
3
MEAS SHAPE
RECTANGLE
4
MEAS AREA
SET (224, 208)
〜
(287, 271)
5
BINRY NOISE FILT
EPD.
→
CONTR.
6
NUM.OF FILT PASS
EPD0 CONT0
(0
〜
5)
7
BOUNDARY PROCESS
VALID
8
AREA FILTER
U. LM245760 L. LM000000
(0
〜
245760)
SET
=SELECT A MENU ESC=BACK SEL=CHNG IMG TRG=FUNC
COLOR F C1 BRT
REG CND
SCREEN COND SAVE DETAIL
■
BINRY NOISE FILT
■
BOUNDARY PROCESS
■
AREA FILTER
RECTANGLE
Specify the coordinates for the upper left and lower right corners.
The measurement area will be the area enclosed by this rectangle.
CIRCLE
Specify the center and the radius. The measurement area will be the
area enclosed by this circle.
ELLIPSE
Specify the center and the radius. The measurement area will be the
area enclosed by this ellipse.