LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
A02 Patient Parameter
and A06 OEM/SpO2
Assembly Removal
Note:
Remove the following assemblies before beginning this disassembly:
–
assembly (optional removal)
To remove the A02 Patient Parameter and optional A06 OEM/SpO2 assemblies:
1. Disconnect the W05 SpO2 Flex Cable from the A06 OEM/SpO2 Assembly at
J54 by removing the screw (255).
2. Disconnect the W06 ECG wire harness from the A02 PP PCB at J23.
3. Remove and discard the 4-40
0.937 screw (125). Remove the three 4-40
nuts (161) from the A02 Patient Parameter PCB.
4. If the device is equipped with the A06 OEM/SpO2 option, remove the Nomex
shield (262) by removing the two washers (264) and nuts (161).
if replacing the A06 OEM PCB.
5. Lift the A02 Patient Parameter PCB (OEM/SpO2) assembly away from the
boardstack assembly. Ensure that the PCB clears the lip on the frame in the
lower right corner.
(Continued on next page)
9-84