LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Using the Capacitor Discharge Tool
(continued)
.
6. Continue holding the probes on the points indicated for at least 30 seconds
after the neon lamp is no longer lit.
WARNING
SHOCK HAZARD.
Do not assume the capacitor is discharged if the neon lamp does
not light! There may still be a charge on the capacitor. Do not touch capacitor
terminals until completing the discharge operation.
Solder point on
inductive resistor
Capacitor discharge
cable connection point
Capacitor discharge cable
passes under the
W01 Therapy Cable and
wraps around the
J13 connector.
Position the therapy cable
ferrites so that they will not
damage the J13 connector
Pin 5 on the A04
Therapy PCB P2
connector. See the
detail.
9-15