LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
Code
Service Code Description
Corrective Action Code
1010
Display update queue error
1. Reload the device software*
2. Replace the A01 System PCB
1013
System detected unexpected UI reset
1. Replace the W18 UI Flex Cable
2. Replace the A05 UI PCB
1014
Voice prompt/audio watchdog failure
1. Reload the device software*
2. Replace the A01 System PCB
1015
USB data time-out error
1. Clear error and perform PIP
2. Replace the W18 UI Flex Cable
1c01
UI FPGA programming error
Replace the A05 UI PCB
1c02
UI FPGA verification error
Replace the A05 UI PCB
1c03
UI FPGA program file error
1. Reload the device software*
2. Replace the A05 UI PCB
1c04
UI ADC not functioning
Replace the A05 UI PCB
(Continued on next page)
6-14