![Physio Control LIFEPAK 20 Service Manual Download Page 299](http://html1.mh-extra.com/html/physio-control/lifepak-20/lifepak-20_service-manual_1553941299.webp)
LIFEPAK 20/20e Defibrillator/Monitor Replacement
Procedures
Bottom Case
(continued)
W05 SpO2 Connector
Removal
Note:
The following assemblies must be removed before beginning this
disassembly:
To remove the W05 SpO2 Connector Assembly:
1. Remove the cable tie retainer (229) securing the SpO2 cable to the bottom
case. The W03 IrDA Port must be removed to expose this tie wrap.
2. Gently pull apart the plastic snap arms on the SpO2 connector mounting clip
(113), away from the bottom case.
3. Lift the mounting clip away from the bottom case.
4. Lift the W05 SpO2 connector assembly away from the bottom case.
9-143