LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Top Case
Top Case Removal
To remove the top case:
1. Disconnect the device from ac power.
2.
from the device.
3. Place the device face down.
4. Remove and discard the two 6-32
0.375 screws (169) securing the bottom
case to the front case.
5. Remove and discard the 6-32
1.75 screws (167) securing the bottom case
to the top case.
6. Place the device on its bottom.
7. Remove and discard the two 6-32
1.75 screws (167) securing the top case
to the front case (outboard screws).
8. Pull the front case slightly away from the top case, and slide the top case up
and away from the rest of the device.
9.
169
167
167