LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
8fff
Additional information related to error code that is
logged before 8fff
No action required for 8fff; refer to error logged
before 8fff.
9004
Unable to initialize therapy control
Replace the A01 System PCB
9005
Defib disabled
Replace the A01 System PCB
9007
Shock not delivered
Reload the device software*
9009
Defib charge time expired
Replace the A07 Battery
900b
Pacing rate out of tolerance
Replace the A01 System PCB
900f
Unable to initialize therapy control
1. Check the stack connector
2. Replace the A04 Therapy PCB
9011
Pacer fault
1. Reload the device software*
2. Replace the A04 Therapy PCB
9017
Pacer disabled
Reload the device software*
901a
Pacer rate storage corrupted
Reload the device software*
6-33