LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Static-Sensitive Devices
(continued)
Wear a Wrist Strap
•
Always wear a conductive wrist strap connected to the mat and to ground
except when working on energized equipment or when discharging high
voltage circuits. The strap must be snug enough to make good contact
against bare skin.
Transport and Store
PCBs Properly
Transport and store PCBs in anti-static racks or inside conductive bags. Label
the package that contains the PCBs as static-sensitive.
Keep Work Area
Static-Free
Keep static-generating products, such as styrofoam cups or trays, away from the
work area. Connect all electrical equipment, such as test equipment, to ground
with a three-prong plug.
Test Work Area
Routinely
Test all antistatic parts of the work area (mat, straps, cables) routinely. Keep a
log of the test results.
WARNING
SHOCK HAZARD.
Remove the wrist strap when working on energized equipment or
when discharging high-voltage circuits.
9-11