LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
5105
Battery failed to reach charge in 2.5 hours
1. Replace the A07 Battery
2. Replace the A03 Power module
5106
Power supply out of tolerance
1. If power supply voltage is <14 Vdc, replace
A03 Power module
2. If >14 Vdc, clear error and perform PIP
5107
Power module RAM error
Replace the A03 Power module
5108
Power module self-test service code
Replace the A03 Power module
510a
Battery <10 V after 20 minutes of charging
1. Replace the A07 Battery
2. Replace the A03 Power module
510b
Does not switch to battery power
Replace the A03 Power module
510c
Does not detect ac disconnect
Replace the A03 Power module
510d
Battery powered when connected to ac power
Replace the A03 Power module
510e
Battery <10 V after charge cycle
Replace the A07 Battery
6-27