LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code Service Code Description
Corrective Action
800a
System DSP error
Reload the device software*
8013
Voice format error
Reload the device software*
8014
No paddles data
1. Reload the device software*
2. Replace the A01 System PCB
801b
DSP did not receive USB SOF (start of frame) interrupt
1. Clear error and perform PIP
2. Replace the A01 System PCB
8105
Impedance channel out of calibration
Replace the A01 System PCB
8108
Paddles data out of sync
1. Reload the device software*
2. Replace the A01 System PCB
8109
Paddles pre-amp user test failed
1. Rerun the user test
2. Replace the A01 System PCB
810b
Real impedance <-30 ohms for one second
1. Perform TCP - defibrillator calibration
2. Replace the A01 System PCB
6-32