LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
9c13
Therapy PCB 15 V out of range
Replace the A04 Therapy PCB
9c14
Therapy PCB -15 V out of range
Replace the A04 Therapy PCB
9c15
Therapy ADC time out error
Replace the A04 Therapy PCB
9c18
3:00 AM H bridge test: NE leg shorted
Replace the A04 Therapy PCB
9c19
3:00 AM H bridge test: SE leg shorted
Replace the A04 Therapy PCB
9c1a
3:00 AM H bridge test: NW leg shorted
Replace the A04 Therapy PCB
9c1b
3:00 AM H bridge test: SW leg shorted
Replace the A04 Therapy PCB
9c1c
3:00 AM H bridge test: east side stuck open
Replace the A04 Therapy PCB
9c1d
3:00 AM H bridge test: west side stuck open
Replace the A04 Therapy PCB
9c1e
3:00 AM H bridge test: charge time out of range
1. Perform TCP - defibrillator calibration
2. Clear Service Log