LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Processing Service Codes
Introduction
When an internal program or process fails to execute properly, a service code is
logged and the service indicator LED turns ON. Service codes rarely occur and
should be investigated thoroughly by qualified service personnel before the
device is placed back into active use. Always complete the performance
inspection procedure (PIP) after encountering and clearing any service code(s).
Service codes stored in the
may not necessarily indicate a
permanent error. Service codes can indicate transient electromagnetic
interference (EMI) or
. If you suspect transient
EMI or ESD as the source of an error, clear the service code(s), and cycle the
power. If the service code does not reoccur, it may have been the result of EMI
or ESD.
Processing Service
Codes
To process service code(s):
1. Note any problems with the device and consult the
2. Review service codes in the
. Record the service code(s),
including the date, time, and service code extension.
3. Select
CLEAR LOG
in the service log, and then turn the device OFF.
(Continued on next page)
6-2