LIFEPAK 20/20e Defibrillator/Monitor
Modes of Operation
Setup Mode
(continued)
Displaying the Setup
Menu
To display the
SETUP
menu:
1. Press and hold the
OPTIONS
and
EVENT
buttons, and then press the
ON
button. When the
ENTER PASSCODE
overlay appears, release the buttons.
The
SPEED DIAL
LED illuminates, indicating that the
SPEED DIAL
is active.
2. To enter the passcode, rotate the
SPEED DIAL
to select a number and then
press the
SPEED DIAL
. As a number is selected, it changes to an asterisk for
passcode protection, and the next digit in line highlights.
Note:
The factory default passcode (
0000
) or the reserved technician
passcode (
5433
or
LIFE
) can be used in place of other passcodes to gain
access to the
SETUP
and
SERVICE
menus.
3. When you have entered the passcode, press the
SPEED DIAL
. The
SETUP
menu appears. The
PASSCODE INCORRECT-TRY AGAIN
message appears if
an incorrect passcode is entered.
To exit the
SETUP
menu, turn the device OFF.
(Continued on next page)
4-6