LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
5115
Battery thermistor <400 Ohms
Replace A07 Battery (LIFEPAK 20e)
Ohm measurement of white wires of A07 Battery
(battery thermistor LIFEPAK 20)
■
Replace the A07 Battery if <400 Ohms
■
Replace the A03 Power module if >400 Ohms
5116
Charger reporting zero during battery charging
Replace the A03 Power module
5117
No 12C connection detected
(LIFEPAK 20e only)
1. Replace the A07 Battery
2. Replace the A03 Power module
5118
Power module connected to NiMH battery
(LIFEPAK 20 only)
Replace the A07 Battery
511a
Battery failure
Replace the A07 Battery
5fff
Additional information related to error code that is
logged before 5fff
No action required for 5fff; refer to error code
logged before 5fff
6-29