![Physio Control LIFEPAK 20 Service Manual Download Page 283](http://html1.mh-extra.com/html/physio-control/lifepak-20/lifepak-20_service-manual_1553941283.webp)
LIFEPAK 20/20e Defibrillator/Monitor Replacement
Procedures
Bottom Case
(continued)
A03 Power Module
Removal
(continued)
4. Pull the boardstack assembly away from the printer shroud and power
module.
5. If replacing the power module,
remove the A13 Energy Capacitor
(complete steps 6 through 9).
6. Remove and discard the forward right 4-40
0.312 screw (173) that secures
the right side of the power module and grounding harness (221).
Note:
Remove the grounding harness (221) as you remove the right forward
screw. Replace the grounding harness if broken or frayed.
7. Remove and discard the forward left 4-40
0.312 screw (173) that secures
the left side of the power module.
8. Loosen the rear 4-40
0.312 screw (173) that secures the right rear corner
of the power bracket three turns.
9. Tilt the left side of the power module up, clearing the loosened screw, and
remove it from the bottom case.