LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Boardstack
(continued)
2. Install the new coin cell battery (344) by inserting the side of the battery
(positive up) against the positive contacts and then pressing it into the
negative side of the battery holder. Ensure that the battery is fully seated in
the holder and that the side of the battery touches the positive contacts
3. After the battery has been replaced, check the voltage on the coin cell
battery holder circuit with a multi-meter to confirm that the battery is installed
correctly and is making contact with the holder.
4. Replace the PCB shield (127).
5.
Install the A02 Patient Parameter PCB and A06 OEM PCB
9-96