LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Front Case
(continued)
A11 Active Display
Removal
(continued)
5. Check the condition of the following parts. Remove and replace any part that
has cracks, broken wires, or damaged connectors.
A08 Backlight Inverter
— To remove, disconnect the Active Display wires
at J2. Disconnect the W17 Backlight Inverter Cable at J1. Pull the backlight
inverter away from the display bracket. A new display bracket is required
because the adhesive and foam are pre-attached.
W15 Active Display Cable
— To remove, disconnect it from the A05 UI
PCB at J36 (cable was previously disconnected from the Active Display in
step 1).
W17 Backlight Inverter Cable
— To remove, disconnect it from the A05 UI
PCB at J37. Disconnect it from the A08 Backlight Inverter (if not previously
removed) at J1. Pull the ferrite bead out of the molded notches on the
display bracket.
Display bracket
(47) — After removing above parts, replace if necessary.
The display bracket has the Backlight Inverter PCB adhesive and the Active
Display foam pre-attached.
Thermal conductive pad
(201) — (used only with 3202033 - A08 Backlight
Inverter) To remove, peel away the old pad located on the top inside edge of
the top case (
).
W17
47
A08
Align to right edge
of bracket flange
illustration 1
9-44