LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Front Case
(continued)
A11 Active Display
Installation
(continued)
3. If replacing the A08 Backlight Inverter, a new display bracket (47) is required.
The display bracket has the adhesive and display foam piece pre-installed.
Remove backing from display bracket. Place the A08 Backlight Inverter on
display bracket so that J1 connector edge is aligned to the top side and right
edge of the bracket flange.
4. Connect the W17 Backlight Inverter Cable “P74” connector to location J1 on
A08 Backlight Inverter.
5. Place W17 Backlight Inverter cable’s ferrite bead into the molded saddle
location such that the cable tie does not interfere with the saddle flange. (
6. Insert the snap rivet (266) through the hole from the back of the display and
ensure the rivet expands on the front side.
7. To install the W15 Active Display cable, unlatch “CN101” connector lock on
A11 Active Display. Insert W15 Active Display cable under latch with metal
contact band facing toward the circuit board until the cable is fully seated and
square in the holding latches on the connector lock to secure the cable.
W15
A11
266
illustration 3
(Continued on next page)
9-46