LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
510f
Battery charge cycle stopped
1. Check battery connection
2. Replace the A07 Battery
3. Replace the A03 Power module
5110
AC isolation diode shorted
Replace the A03 Power module
5112
Battery not detected
1. Check the A07 Battery connection
2. Check W08 Battery Cable
3. Check for valid power hardware ID in
SERVICE/STATUS/DEVICE
DATA
overlay
4. Replace the A07 Battery
5. Invalid power hardware ID; replace the
A03 Power module
6. Valid power hardware ID; replace the
A04 Therapy PCB
6-28