LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
6801
PP power supply out of tolerance
Replace the A02 PP PCB
6802
PP pre-amp data invalid
1. Clear error and perform PIP
2. Replace the A02 PP PCB
6804
PP data RAM test error
1. Clear error and perform PIP
2. Replace the A02 PP PCB
6805
PP RAM test error
1. Clear error and perform PIP
2. Replace the A02 PP PCB
6806
PP CRC test error
1. Clear error and perform PIP
2. Replace the A02 PP PCB
6807
PP ECG test error
1. Clear error and perform PIP
2. Replace the A02 PP PCB
680b
SpO2 board error
Replace the A10 SpO2 Module
6fff
Additional information related to error code that is logged
before 6fff
No action required for 6fff; refer to error logged
before 6fff.
6-31