LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Front Case
(continued)
AED Door
Replacement
The AED door assembly is designed to be an easily replaceable, breakaway
assembly. If the door assembly accidently comes off during use, follow step 3 of
the AED door installation procedure below.
To remove the entire AED door assembly:
1. Open the AED door.
2. Use a small screwdriver to pry the hinge pin center slightly away from the
door assembly until the door slides free of the hinge.
3. Peel the hinge off the front case.
4. Clean the front case to remove old adhesive.
To install a new AED door assembly:
1. Clean the hinge area.
2. Expose the adhesive and secure the door hinge plate (37) to the front case.
3. Use a small screwdriver to pry the hinge pin center slightly away from the
door assembly until the door slides into the hinge. Ensure that the hinge pins
snap into the securing holes.
9-39