LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Front Case
(continued)
A11 Active Display
Installation
(continued)
14. Install four new 4-40
0.312 screws (173) to secure the display assembly to
the front case; torque to 6.8 in-lb.
15. Connect the W15 Active Display Cable to the A05 UI PCB at J36 (if removed
previously), as follows.
–
Open the J36 connector lock.
–
Insert the W15 Display Cable (metal contacts down) into the connector
lock.
–
Close the connector lock to secure the cable.
Note:
The cable connector must be square with the connector lock.
16. Complete the process by
.
CAUTION
Possible component damage.
The grounding harnesses must be installed at
precise angles to avoid damaging device components.
9-48