![Physio Control LIFEPAK 20 Service Manual Download Page 85](http://html1.mh-extra.com/html/physio-control/lifepak-20/lifepak-20_service-manual_1553941085.webp)
LIFEPAK 20/20e Defibrillator/Monitor
Modes of Operation
Setup Mode
(continued)
Creating a Passcode
■
ARCHIVES ACCESS
– Set the device to any of the following protocols (refer to
the table above):
1. Allow unlimited access to archives mode and allow records to be deleted.
2. Require a password to enter archives mode, but allow records to be
deleted.
3. Allow unlimited access to archives mode, but require a password to delete
records.
4. Require a password to enter archives mode and delete records setup
MODE
– Create a new passcode to access the
SETUP
menu. The
ENTER
PASSCODE
overlay appears with the first digit highlighted. Rotate the
SPEED DIAL
to select digits.
■
ARCHIVES MODE
– Create a passcode to enter archives mode. The
ENTER
PASSCODE
overlay appears with the first digit highlighted. Rotate the
SPEED
DIAL
to select digits.
■
DELETE RECORDS
– Create a passcode to delete records in archives mode.
The
ENTER PASSCODE
overlay appears with the first digit highlighted. Rotate
the
SPEED DIAL
to select digits.
4-10