LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
1c11
UI application Flash memory corrupted
1. Reload the device software*
2. Replace the A05 UI PCB
1c12
UI font Flash memory corrupted
1. Reload the device software*
2. Replace the A05 UI PCB
1c13
UI FPGA Flash memory corrupted
1. Reload the device software*
2. Replace the A05 UI PCB
1c16
UI CPU RAM test failed on power-on
Replace the A05 UI PCB
1c17
UI CPU RAM test failed during normal operation
Replace the A05 UI PCB
1fff
Additional information related to error code that is
logged before 1fff
No action required; refer to error code logged
before 1fff.
2004
System cannot initialize serial port (system connector)
Replace the A01 System PCB
2005
System cannot initialize driver for serial port (system
connector)
1. Reload the device software*
2. Replace the A01 System PCB