LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
Code
Service Code Description
Corrective Action
1c05
Printer ADC out of tolerance (temperature)
1. Check the W14 Printer Flex Cable
2. Replace the A12 Printer
3. Replace the A05 UI PCB
1c06
Printer ADC out of tolerance (voltage)
1. Check the W14 Printer Flex Cable
2. Replace the A12 Printer
3. Replace the A05 UI PCB
1c07
UI voltage out of tolerance (5 V)
Replace the A05 UI PCB
1c08
UI voltage out of tolerance (3.3 V)
Replace the A05 UI PCB
1c09
UI voltage out of tolerance (2.5 V)
Replace the A05 UI PCB
1c0a
UI voltage out of tolerance (35 V)
Replace the A05 UI PCB
1c0b
UI voltage out of tolerance (SW VBATT)
Replace the A05 UI PCB
1c0c
UI voltage out of tolerance (ground)
Replace the A05 UI PCB
1c0f
UI hardware I.D. corrupted
Replace the A05 UI PCB
1c10
UI boot program corrupted
Replace the A05 UI PCB
(Continued on next page)
6-15