LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
6002
PP program corrupted
Reload the device software*
6003
PP program not found
Reload the device software*
6004
PP boot-up error
Reload the device software*
6009
No PP data
Reload the device software*
600c
SpO2 misconfigured
1. Check configuration
2. Reload the device software*
600e
PP reset
1. Reload the device software*
2. Replace the A02 PP PCB
600f
OEM configuration error
1. Check configuration
2. Reload the device software*
6010
PP initialization error
Reload the device software*
6018
Incorrect updated SpO2 image
Replace the A06 OEM PCB
6019
Incorrect updated SpO2 image CRC
Replace the A06 OEM PCB
6-30