LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Front Case
(continued)
W18 UI Flex Cable
Removal
Note:
must be removed before beginning this disassembly.
To remove the W18 UI Flex Cable:
1. From the system shield, disconnect the two grounding straps (219) that
connect the top of the front case to the system shield, by removing the two
screws.
Note:
Observe the
positioning on the grounding harnesses
not reinstalled at the correct angles, the front case will not join with the rest of
the device correctly.
2. Pull the front case slightly forward, away from the boardstack assembly, and
disconnect the W18 UI Flex Cable from the A02 PP PCB at J21 and J22.
3. Disconnect the W18 UI Flex Cable from the A05 UI PCB at J31, and remove
the cable from the device.
9-40