LIFEPAK 20/20e Defibrillator/Monitor Replacement
Procedures
Bottom Case
(continued)
W03 IrDA Assembly
Installation
To install the W03 IrDA Assembly:
1. Position the W03 IrDA Assembly on the bottom case. (The IrDA is located
near the center of the front panel on the bottom case.)
2. Insert a large slotted screwdriver into the slot between the bottom case
cutout (A) and the right snap tab (B).
3. Gently apply pressure to the screwdriver. Bend the right IrDA snap tab (B)
slightly outward,
4. Press the W03 IrDA Assembly down into the snap tabs and release the
pressure on the screwdriver. The snap tabs will close around the W03 IrDA
Assembly.
5. Ensure that the W03 IrDA Assembly is resting centered on the support
brackets and snap tabs.
6. Connect the W03 IrDA Assembly to the A01 System PCB at J08 and fasten
with a screw (255) using a 3/32 allen driver, if the boardstack is still installed
in the bottom case. Otherwise,
. The IrDA flex cable
is connected during the boardstack installation.
B
A
(Continued on next page)
9-147