LIFEPAK 20/20e Defibrillator/Monitor Replacement
Procedures
Bottom Case
(continued)
W01 Therapy
Connector Installation
To install the W01 Therapy Connector Assembly:
1. From outside the case, align the key on the connector with the notch in the
bottom case and slide the W01 Therapy Connector Assembly into the
bottom case.
2. Install the retaining ring (111) onto the back of the W01 Therapy Connector
Assembly, and rotate the clip so that the open end is visible.
3. Install the Therapy Connector Seal (268) onto the W01 Therapy Connector.
4.
5.
6.
.
7.
8. Complete the LIFEPAK 20 Device PIP.