LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
5029
Therapy USB did not initialize
1. Clear error and perform PIP
2. Replace the A04 Therapy PCB
502a
UI USB did not initialize
1. Clear error and perform PIP
2. Replace the W18 UI Flex Cable
3. Replace the A05 UI PCB
502b
USB system failed
1. Clear error and perform PIP
2. Replace the A01 System PCB
502f
Device type invalid
1. Check configuration code
2. Set device type (software loader program)
3. Replace the A01 System PCB
5030
PP USB disconnect
1. Clear error and perform PIP
2. Replace the A02 PP PCB
5031
UI USB disconnect
1. Clear error and perform PIP
2. Replace the W18 UI Flex Cable
3. Replace the A05 UI PCB
6-24