LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Saving the Setup Configuration
The following procedures describe how to save the device setup configuration
before beginning any repair action.
■
If the device owner has a LIFENET System account, then use the LIFENET
Device Agent application to manage setup configurations.
■
Another method is to transfer the setup configuration to a spare device,
complete repairs, and then transfer the setup configuration back again.
■
The final method is to print the setup configuration, complete repairs, and
then manually reconfigure the device.
Note:
Saving the configuration by transferring it to a spare device requires that
both devices have the same software version. Otherwise, potentially
unexpected results may occur when the configuration is restored to the
repaired device. Verify that copyright dates are the same on the
introduction page of both devices.
9-16