LIFEPAK 20/20e Defibrillator/Monitor
Modes of Operation
Service Mode
(continued)
Displaying the Service
Menu
To display the
SERVICE
menu:
1. Display the
2. Select
SERVICE
from the
SETUP
menu.
3. Enter the service mode passcode (
0000
or
5433
).
4. After you enter the passcode, press the
SPEED DIAL
. The
SERVICE
menu
appears. (If an incorrect passcode is entered, the
PASSCODE INCORRECT -
TRY AGAIN
message appears.)
Service Menu Options
The
SERVICE
menu options include:
To exit the
SERVICE
menu, turn the device OFF.
Option
Description
Perform defibrillator calibration procedure.
Follow performance inspection procedure.
Display device status.
Set the service mode access passcode.
Prompt user to perform preventative maintenance.
Setup
Return to main
SETUP
menu.
(Continued on next page)
4-12