LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
Code
Service Code Description
Corrective Action
3001
System cannot use data management (DM) Flash
memory
1. Reload the device software*
2. Replace the A01 System PCB
3002
System data management Flash memory corrupted
1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
3005
System cannot delete DM record
1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
3007
System cannot create new DM record
1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
3008
System could not store DM record
1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
(Continued on next page)
6-17