LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
Code
Service Code Description
Corrective Action
a00b
Printer communication lost
1. Clear error and perform PIP
2. Reload the device software*
a00e
Printer initialization error
Reload the device software*
b001
Invalid state request
1. Clear error and perform PIP
2. Replace A01 System PCB
b011
System behavior manager error
Reload the device software*
b012
Energy cap charging time out
Reload the device software*
b013
Shock advisory system error
Reload the device software*
b014
Shock advisory system time out
Reload the device software*
b016
Motion detect timer error
Reload the device software*
b017
Shock result time out
Reload the device software*
b018
USB interrupt error
1. Clear error and perform PIP.
2. Replace A01 System PCB.
(Continued on next page)
6-40