LIFEPAK 20/20e Defibrillator/Monitor
Modes of Operation
Service Mode
(continued)
Setting a Maintenance
Prompt Interval
The LIFEPAK 20 defibrillator/monitor can be set to display a screen message
that alerts the user when the maintenance prompt interval date has passed. The
screen message
MAINTENANCE DUE
appears on the screen for the first
10 minutes after the device is powered on. The device maintenance interval can
be turned off or set to 3 months, 6 months, or 12 months; the factory default is
OFF,
but it can be activated by a service technician.
To change the scheduled maintenance interval:
1. Display the
SERVICE
menu.
2. Select
MAINT PROMPT
. The
SERVICE/MAINT PROMPT
submenu appears
showing the current prompt date for scheduled maintenance (if set).
3. Select
INTERVAL
. The interval choices are:
OFF
,
3 MONTHS
,
6 MONTHS
, and
12 MONTHS
.
4. Select the desired interval.
5. Turn the device OFF.
(Continued on next page)