LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Front Case
(continued)
Front Case Removal
(continued)
6. Disconnect the two grounding harnesses (221) that connect the bottom left
and right corners of the A05 User Interface (UI) PCB to the PCB support
bracket by removing and discarding the two screws (173).
7. Disconnect the grounding harness (219) that connects the bottom center of
the A05 UI PCB to the PCB support bracket by removing and discarding the
screw (173).
8. Disconnect the W25 Speaker Harness Extension Cable from the W02
Speaker Assembly.
9. Disconnect the W14 Printer Flex Cable from the A05 UI PCB at J34.
Note:
Disconnect the Speed Dial connector to access the printer connector.
10. Pull the front case away from the device.
9-34