LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
Code
Service Code Description
Corrective Action
5011
NVRAM configuration data error
1. Replace the coin cell battery
2. Reload the device software*
5012
Configuration data error
Replace the coin cell battery
5013
System meter initialization error
Reload the device software*
5014
System meter mismatch
Replace the coin cell battery
5015
NVRAM error
Replace the A01 System PCB
5016
MFG data mismatch
Replace the A01 System PCB
5017
NVRAM MFG data lost
Replace the A01 System PCB
5018
Watchdog reset failed
Replace the A01 System PCB
5019
NVRAM corrupted
Replace the coin cell battery
501a
NVRAM corrupted
Replace the coin cell battery
501c
RTC not running
1. Replace the coin cell battery
2. Replace the A01 System PCB
(Continued on next page)
6-22