LIFEPAK 20/20e Defibrillator/Monitor
Replacement Procedures
Front Case
(continued)
A11 Active Display
Removal
Note:
Remove the following assemblies before beginning this disassembly:
To remove the A11 Active Display:
1. Disconnect the W15 Active Display Cable (see illustration) from the
A11 Active Display, as follows:
–
Gently pull both sides of the locking tab away from the connector.
–
Pull the cable out of the socket (leave the cable connected to the UI
PCB).
2. Disconnect the W17 Backlight Inverter Cable from the A05 UI PCB at J37.
3. Remove and discard the four 4-40
0.312 screws (173) from the display
assembly cover.
Remove the two grounding harnesses
(219) attached to the top
two screws of the display assembly. Replace any broken or frayed grounding
harnesses.
4. Remove the display bracket assembly from the front case.
W15
(connects to A05)
A11
W17
A08
47
173
219
(x4)
(x2)
(Continued on next page)
9-43