LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Troubleshooting Chart
Area
Observed Symptom
Suggested Corrective Action
Physical Inspection
Loose or broken hardware
Locate and tighten or replace loose items.
Locate and replace broken components.
Evidence of dirt, fluids, or
foreign objects
Perform
Damaged keypad or labels
.
Replace product identification label.
Replace explosion/hazard label.
Replace operating instruction label.
Power Off
Device beeps when turned
off
Connect device to ac power source.
Disable
.
(Continued on next page)
6-4