LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
9c1f
3:00 AM shorted paddles relay contact test: relay
shorted
1. Inductive resister is not connected or is open
2. Replace the A04 Therapy PCB
3. Replace the A01 System PCB
9c20
3:00 AM shorted paddles relay contact test: relay
shorted
Replace the A04 Therapy PCB
9c21
3:00 AM pace drive test: pace power supply stuck on
Replace the A04 Therapy PCB
9c22
3:00 AM pace drive test: pace power supply inoperable Replace the A04 Therapy PCB
9c23
3:00 AM pace drive test: relay contacts shorted
Replace the A04 Therapy PCB
9c24
3:00 AM pace drive test: relay drive low side shorted
Replace the A04 Therapy PCB
9c25
3:00 AM pace drive test: relay drive high side shorted
Replace the A04 Therapy PCB
9c26
3:00 AM pace drive test: pace FET shorted
Replace the A04 Therapy PCB
9c27
3:00 AM pace drive test: pace current path open
Replace the A04 Therapy PCB
9c28
3:00 AM pace drive test: pace set point error
Replace the A04 Therapy PCB
6-37