4 - 4
Test settings
Figure 3:
CTRatio
test card with
test results
Navigate to the parameter fields, and enter the values according to your test
requirements:
Pressing the S
E T T I N G S
menu key opens the
Settings
page. The
Settings
page allows setting the test cards individually. As a rule, do not use the
Settings
page but the
Device Setup
tab in the
Options
view (see ”Device Setup” on
page 2-33) to set the test cards. For more information, see ”Settings Page” on
page 2-28.
Range:
output range
Iprim:
nominal primary current
Isec:
nominal secondary current
Itest:
primary injection current.
f:
output frequency
Auto
select check box for automatic test (default), clear for manual
test (refer to page 4-7).
Note:
If the test involves either an external booster to gain a higher
output current or a current clamp to measure Isec, set these
devices at the
Options
tab
Device Setup
before starting the test.
Measure
Burden
The option "Measure Burden" lets you measure the burden, the
secondary voltage and the cos
ϕ
(cosinus of angle between
secondary current and secondary voltage).
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...