Technical Data
16 - 23
300V V1 AC Input
Table 16:
300V V1AC
input
technical data
• DC operation
(autoranging)
10A:
typ. error < ± 0.03% (rd) ± 0.08% (fs),
guaranteed < ± 0.05% (rd) ± 0.15% (fs)
1A:
typ. error < ± 0.03% (rd) ± 0.08% (fs),
guaranteed < ± 0.05% (rd) ± 0.15% (fs)
Input resistance
< 100m
Ω
Resolution
Measurement
Display
10A:
1 mA/bit
1A:
100
μ
A/bit
10
μ
A
Fuse
Type: 10A FF
The 10 Amps very quick-acting wire fuse 5x20mm,
situated at the front panel’s INPUT group, protects the
10A IAC/DC current measuring input.
Make sure the fuse you use is a
10A FF
fuse. Do not
use any other type of fuse!
Note:
Do not connect the 2kV output to this input! The 2kV output is
measured internally.
10A IAC/DC input
300V V1AC input
Connection
4mm banana sockets
Measurement
ranges (autoranging)
0 ... 300V / 0 ... 30V / 0 ... 3V / 0 ... 300mV