CPC 100 V1.41
16 - 16
Table 9:
Technical data of
6A / 130V AC
output in
130V AC mode
Figure 4:
Typical maximum
output voltage of
6A / 130VAC
output in
130V AC mode
depending on frequency
6A / 130V AC output in 130V AC mode
Output power at:
typical apparent
power S
max
typical time t
max
frequency
130V
390VA
> 2hrs
50 ... 400Hz
Fuses
Type: 3.15A T
The 3.15 Amps slow-acting wire fuse 5x20mm,
situated at the front panel’s
AC OUTPUT
group,
protects the
AC OUTPUT
in 130V voltage mode.
6.3A T
If the 6.3 Amps fuse is blown, the 130V AC output
mode does not work either.
Note:
The 130V AC output has a setting range of 150V. The set voltage
represents the no-load voltage, that is, when a load is connected, the
voltage at the 130V output will reduce by approx. 5V / A.
volt
age V
rms
frequency in Hz