CPC 100 V1.41
2 - 48
9. Start the test by pressing the green I/O (test start/stop) push-button on
CPC 100
front panel.
10.If a potentially hazardous voltage and/or current level is applied to the
CPC 100
outputs, the red signal red light "I" starts flashing.
11.After testing, switch off the high voltage or current immediately with the I/O
(test start/stop) push-button.
12.If you do not use
CPC 100
anymore, turn the safety key to "lock" (vertical)
and remove the key to avoid anybody accidentally turning on the high voltage
or current.
After
turning off the high voltage or current observe the warning lights: only
when
CPC 100
green warning light is on and the red one off, switch off the
safety key lock and take off its key.
13.After turning off
CPC 100
outputs, earth-connect and short-circuit the test
object’s terminals again using a grounding set.
14.Remove the connection between
CPC 100
and the test object.
Never touch any metallic terminals and/housing components without a visible
ground connection!
Notes:
• A test can only be started in the Test Card View and with all
preconditions met (refer to ”Starting a Test” on page 29 of this
chapter).
• If voltages
≥
1kV are to be applied, after pressing the I/O
(test start/stop) push-button for the first time a warning
message appears on the screen.
Only after the I/O push-button is pressed a second time, the
voltage is applied to the
CPC 100
output.
• An error message (313) appears if either the PE connection is
defective or the power supply has no galvanic connection to
ground. In this case, make sure that the PE connection is intact.
If the PE connection is intact and the error message still
appears, select the "Disable ground check" check box at the
Device Setup
tab in the
Options
view (see ”Device Setup” on
Rather than starting a test and activating the outputs, pressing
the I/O (test start/stop) push-button will then pop up a message
box notifying you about this problem.
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...